APS122
Overview
Highly accurate with fast response times, the 100 GP-D provides complete confidence in a wide range of general process applications including drinking water, municipal wastewater and cooling waters.
The 100 GP-D sensor provides:
EZLink™ connectivity
- Plug-and-play technology makes sensor integration fast and easy
- Advanced diagnostics offering end-of-life indication and fault analysis
- Improved measurement accuracy with digital communication
Increased efficiency
- Propriety high performance glassware formulations provide the highest level of reliability and responsiveness
- Close-coupled temperature measurement ensures high accuracy even with rapid temperature changes
Cost-effective and dependable
- Enhanced double junction design with ion trap delays poisoning effects ensuring the sensor operates longer
- Large surface PTFE junction provides excellent stability and fouling resistance
- Robust PVDF ¾ inch NPT body provides superior mechanical and chemical resistance
- Each electrode is manufactured to the highest quality standards and provided with a calibration certificate
Modular design
- Universal ¾ inch NPT design for simplified mounting
- Available in a range of detached or integral cabling options
Data
Measurement range
pH 0 to 14 pH
Redox (ORP) –2000 to 2000 mV
Process temperature
–5 to 60 °C (23 to 140 °F)
Pressure range
0 to 6 bar (90 psi) @ 25 °C (77 °F)
Recommended minimum sample conductivity
50 μS/cm
pH glass
High performance (S) glass (bullet)
High performance (S) glass (flat)
Low temperature (LT) glass
Redox (ORP)
Platinum
Reference system
Extended double junction Ag/AgCl design plus ion trap in gel matrix
Reference junction
Porous PTFE and Viton O-rings
Connection type
EZLink
Process connections
¾ inch NPT
Approvals, certification and safety
- CE Mark
- Covers EMC+LV directives (including latest version of EN61010)
- Regulation 31
- Drinking water approval: Complies to DWI Regulation 31(4)(b)
- Additional tests: BS6920 parts 2.2 and 2.4 on all wetted parts
- EMC
- Meets requirements of IEC61326 for an industrial environment
- CRN Approval
- Maximum allowable working pressure (MAWP): 5.58 bar (81 psi)
- Design temperature: -5 ⁰C to + 105 ⁰C (+23 ⁰F to +221 ⁰F)
- CRN number 0F22557
- MCERTS
- Certificate no. Sira MC220375/00